Human GM-CSF Ab(Anti-Granulocyte Macrophage Colony Stimulating Factor Antibody) ELISA Kit

This kit is based on indirect ELISA detection method and takes 3h assay time. The microplate provided in this kit has been precoated with antigen. HRP-antibody is used as detection antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add HRP-detection antibody, then it binds with GM-CSF Ab bound to precoated antibody. Wash unbound components and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of GM-CSF Ab in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
EH3153
Alias: 
GM-CSF antibody ELISA Kit, Granulocyte-macrophage colony-stimulating factor antibody ELISA Kit
Species: 
Human
Range: 
0.625-40ng/ml
Sensitivity: 
0.375ng/ml
  • SPECIFICATIONS
Product Name
Human GM-CSF Ab(Anti-Granulocyte Macrophage Colony Stimulating Factor Antibody) ELISA Kit
Alias
GM-CSF antibody ELISA Kit, Granulocyte-macrophage colony-stimulating factor antibody ELISA Kit
Catalogue No.
EH3153
Size
48T/96T
Species
Human
UniProt No.
P04141
Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method
Indirect ELISA
Detection Wavelength
OD450
Reaction Duration
3 hours
Range
0.625-40ng/ml
Sensitivity
0.375ng/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with GM-CSF Ab , no obvious cross reaction with other analogues.
ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 HRP Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Assay Procedure Summary
Elisa实验原理图
  • Step 1: Wash the plate twice before adding the standard and sample.
  • Step 2: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-antibody working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(ng/ml) OD-1 OD-2 Average
0 0.08 0.084 0.082
0.625 0.128 0.135 0.131
1.25 0.191 0.201 0.195
2.5 0.298 0.313 0.304
5 0.562 0.59 0.573
10 0.927 0.974 0.946
20 1.455 1.53 1.485
40 2.044 2.149 2.086
EH3153 Standard Curve Image
Recovery

Add a certain amount of GM-CSF Ab into the sample. Calculate the recovery by comparing the measured value with the expected amount of GM-CSF Ab in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 85-98 98
EDTA plasma(n=10) 91-102 93
Heparin plasma(n=10) 91-99 94
Linearity

Dilute the sample with a certain amount of GM-CSF Ab at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 88-103% 89-103% 100-105%
EDTA plasma(n=10) 89-99% 85-99% 85-95%
Heparin plasma(n=10) 82-100% 84-96% 86-96%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/ml) 1.17 5.3 19.1 1.31 5.09 20.41
Standard deviation 0.07 0.28 1.03 0.08 0.3 0.86
CV(%) 5.91 5.23 5.39 6.43 5.92 4.23
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100