HRP-SA(Streptavidin Horseradish Peroxidase)Activity ELISA Kit

Category:
Catalogue No.: 
EU2619
Alias: 
HRP-SA ELISA Kit, Streptavidin Horseradish Peroxidase ELISA Kit
Species: 
Universal
Range: 
28.125-1800uIU/ml(1800uU=6ng)
Sensitivity: 
14uIU/ml
  • SPECIFICATIONS
Product Name
HRP-SA(Streptavidin Horseradish Peroxidase)Activity ELISA Kit
Alias
HRP-SA ELISA Kit, Streptavidin Horseradish Peroxidase ELISA Kit
Catalogue No.
EU2619
Size
48T/96T
Species
Universal
Sample Type
Quantitative?detection?of?HRP-SA?(horseradish?peroxidase?conjugated?streptavidin)
Detection Method
Direct ELISA
Detection Wavelength
OD450
Reaction Duration
1 hours
Range
28.125-1800uIU/ml(1800uU=6ng)
Sensitivity
14uIU/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with SA-HRP , no obvious cross reaction with other analogues.
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(uIU/ml) OD-1 OD-2 Average
0 0.153 0.157 0.155
28.125 0.191 0.197 0.194
56.25 0.281 0.289 0.285
112.5 0.419 0.431 0.425
225 0.597 0.615 0.606
450 0.909 0.935 0.922
900 1.606 1.652 1.629
1800 2.21 2.274 2.242
EU2619 Standard Curve Image
Recovery

Add a certain amount of SA-HRP into the sample. Calculate the recovery by comparing the measured value with the expected amount of SA-HRP in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 88-101 96
EDTA plasma(n=10) 92-104 98
Heparin plasma(n=10) 87-99 95
Linearity

Dilute the sample with a certain amount of SA-HRP at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 87-105% 87-97% 85-100%
EDTA plasma(n=10) 93-100% 83-98% 85-98%
Heparin plasma(n=10) 88-98% 82-94% 82-96%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (uIU/ml) 57.23 222 892 56.58 226 878
Standard deviation 2.32 10.41 55.57 2.95 12.93 50.22
CV(%) 4.06 4.69 6.23 5.22 5.72 5.72
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100