Products
- SPECIFICATIONS
- Product Name
- POD(peroxidase) Assay kit
- Alias
- POD ELISA Kit, peroxidase ELISA Kit
- Catalogue No.
- EU2611
- Size
- 48T/96T
- Species
- Universal
- Sample Type
- Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
- Detection Method
- Enzyme catalysis
- Detection Wavelength
- OD450
- Reaction Duration
- 20 minutes
- Range
- 3.12-200pmol/ml
- Sensitivity
- 1.875pg/ml
- Storage
- 2-8°C(Sealed), Don't cryopreserve.
- Specificity
- Specifically binds with POD , no obvious cross reaction with other analogues.
- Required Instruments and Reagents
-
- Microplate reader (wavelength: 450nm)
- 37°C incubator (CO2 incubator for cell culture is not recommenced.)
- Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
- Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
- Sterile tubes and Eppendorf tubes with disposable tips
- Absorbent paper and loading slot
- Deionized or distilled water
- Standard Curve
-
This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)
Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.
STD.(pmol/ml) OD-1 OD-2 Average 0 0.07 0.066 0.068 3.125 0.275 0.23 0.252 6.25 0.333 0.344 0.338 12.5 0.582 0.609 0.595 25 0.817 0.82 0.818 50 1.412 1.435 1.423 100 1.81 1.892 1.851 200 2.692 2.615 2.653 - Recovery
-
Add a certain amount of POD into the sample. Calculate the recovery by comparing the measured value with the expected amount of POD in the sample.
Sample Type Recovery Range(%) Average(%) serum(n=10) 91-104 99 EDTA plasma(n=10) 90-104 96 Heparin plasma(n=10) 95-105 99 - Linearity
-
Dilute the sample with a certain amount of POD at 1:2, 1:4 and 1:8 to get the recovery range.
Sample Type 1:2 1:4 1:8 serum(n=10) 90-98% 88-100% 85-100% EDTA plasma(n=10) 84-99% 83-99% 82-96% Heparin plasma(n=10) 83-93% 80-93% 86-100% - Precision(%)
-
Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.
Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.
Item Intra-assay Precision Inter-assay Precision Sample 1 2 3 1 2 3 n 20 20 20 20 20 20 Mean (pmol/ml) 5.88 23.69 102.4 6 23.85 101.41 Standard deviation 0.31 1.21 5.56 0.28 1.49 5.05 CV(%) 5.32 5.12 5.43 4.69 6.23 4.98 - Stability
-
Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.
ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months Average(%) 80 95-100