Monkey NSE(Neuron-specific Enolase) ELISA Kit

Catalogue No.: 
EMK0214
Alias: 
NSE ELISA Kit, ENO2 ELISA Kit, gamma-Enolase ELISA Kit, enolase 2(gamma, neuronal) ELISA Kit, gamma-enolase ELISA Kit, Neural enolase ELISA Kit, neuron specific gamma enolase ELISA Kit, Neuron-specific enolase ELISA Kit
Species: 
Monkey
Range: 
0.313-20ng/ml
Sensitivity: 
0.188ng/ml
  • SPECIFICATIONS
Product Name
Monkey NSE(Neuron-specific Enolase) ELISA Kit
Alias
NSE ELISA Kit, ENO2 ELISA Kit, gamma-Enolase ELISA Kit, enolase 2(gamma, neuronal) ELISA Kit, gamma-enolase ELISA Kit, Neural enolase ELISA Kit, neuron specific gamma enolase ELISA Kit, Neuron-specific enolase ELISA Kit
Catalogue No.
EMK0214
Size
48T/96T
Species
Monkey
Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method
Sandwich ELISA, Double Antibody
Detection Wavelength
OD450
Reaction Duration
4 hours
Range
0.313-20ng/ml
Sensitivity
0.188ng/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with NSE , no obvious cross reaction with other analogues.
ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 Biotin-labeled Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E034 HRP-Streptavidin Conjugate(SABC, 100X) 60ul 120ul
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E049 SABC Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(ng/ml) OD-1 OD-2 Average
0 0.112 0.116 0.114
0.312 0.152 0.156 0.154
0.625 0.196 0.202 0.199
1.25 0.293 0.301 0.297
2.5 0.526 0.542 0.534
5 0.865 0.891 0.878
10 1.507 1.551 1.529
20 2.163 2.225 2.194
EMK0214 Standard Curve Image
Recovery

Add a certain amount of NSE into the sample. Calculate the recovery by comparing the measured value with the expected amount of NSE in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 86-102 93
EDTA plasma(n=10) 88-96 92
Heparin plasma(n=10) 88-102 94
Linearity

Dilute the sample with a certain amount of NSE at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 87-99% 85-95% 85-100%
EDTA plasma(n=10) 88-97% 94-99% 89-96%
Heparin plasma(n=10) 89-101% 85-93% 86-100%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/ml) 0.59 2.54 9.53 0.61 2.55 9.84
Standard deviation 0.03 0.11 0.51 0.04 0.14 0.56
CV(%) 5.34 4.23 5.39 5.85 5.42 5.72
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100