Products
Human CMLC-1(Cardiac Myosin Light Chains 1) ELISA Kit
This kit is based on Double antibody-Sandwich ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with anti CMLC-1 antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antibody. Then, it binds with CMLC-1 bound to precoated antibody. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of CMLC-1 in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.
- Catalogue No.:
- EH2845
- Alias:
- Myosin light chain 3 ELISA Kit, Cardiac myosin light chain 1 ELISA Kit, CMLC1 ELISA Kit, Myosin light chain 1, slow-twitch muscle B/ventricular isoform ELISA Kit, MLC1SB ELISA Kit, Ventricular myosin alkali light chain ELISA Kit, Ventricular myosin light chain 1 ELISA Kit, VLCl ELISA Kit, Ventricular/slow twitch myosin alkali light chain ELISA Kit, MLC-lV/sb ELISA Kit, MYL3 ELISA Kit
- Species:
- Human
- Range:
- 0.625-40ng/ml
- Sensitivity:
- 0.375ng/ml
- SPECIFICATIONS
- FIGURES
- CONDITIONS
- Product Name
- Human CMLC-1(Cardiac Myosin Light Chains 1) ELISA Kit
- Alias
- Myosin light chain 3 ELISA Kit, Cardiac myosin light chain 1 ELISA Kit, CMLC1 ELISA Kit, Myosin light chain 1, slow-twitch muscle B/ventricular isoform ELISA Kit, MLC1SB ELISA Kit, Ventricular myosin alkali light chain ELISA Kit, Ventricular myosin light chain 1 ELISA Kit, VLCl ELISA Kit, Ventricular/slow twitch myosin alkali light chain ELISA Kit, MLC-lV/sb ELISA Kit, MYL3 ELISA Kit
- Catalogue No.
- EH2845
- Size
- 48T/96T
- Species
- Human
- UniProt ID
- P08590
- Sample Type
- Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
- Detection Method
- Sandwich ELISA, Double Antibody
- Detection Wavelength
- OD450
- Reaction Duration
- 4 hours
- Range
- 0.625-40ng/ml
- Sensitivity
- 0.375ng/ml
- Storage
- 2-8°C(Sealed), Don't cryopreserve.
- Specificity
- Specifically binds with CMLC-1 , no obvious cross reaction with other analogues.
- ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C E003 Biotin-labeled Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light) E034 HRP-Streptavidin Conjugate(SABC, 100X) 60ul 120ul E024 TMB Substrate 5ml 10ml E039 Sample Dilution Buffer 10ml 20ml 2-8°C E040 Antibody Dilution Buffer 5ml 10ml E049 SABC Dilution Buffer 5ml 10ml E026 Stop Solution 5ml 10ml E038 Wash Buffer(Concentrated, 25X) 15ml 30ml E006 Plate Sealer 3 pieces 5 pieces E007 Product Description 1 copy 1 copy - Required Instruments and Reagents
-
- Microplate reader (wavelength: 450nm)
- 37°C incubator (CO2 incubator for cell culture is not recommenced.)
- Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
- Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
- Sterile tubes and Eppendorf tubes with disposable tips
- Absorbent paper and loading slot
- Deionized or distilled water
- Assay Procedure Summary
-
- Step 1: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
- Washing: Wash the plate twice without immersing.
- Step 2: Add 100ul biotin-antibody working solution, seal the plate and statically incubate for 60 minutes at 37°C.
- Washing: Wash the plate three times and immerse for 1min each time.
- Step 3: Add 100ul HRP-Streptavidin Conjugate (SABC) working solution, seal the plate and statically incubate for 30 minutes at 37°C.
- Washing: Wash the plate five times and immerse for 1min each time.
- Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
- Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.
- Standard Curve
-
This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)
Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.
STD.(ng/ml) OD-1 OD-2 Average 0 0.072 0.074 0.073 0.625 0.091 0.093 0.092 1.25 0.12 0.124 0.122 2.5 0.207 0.213 0.21 5 0.341 0.351 0.346 10 0.774 0.796 0.785 20 1.357 1.397 1.377 40 2.142 2.204 2.173 - Recovery
-
Add a certain amount of CMLC-1 into the sample. Calculate the recovery by comparing the measured value with the expected amount of CMLC-1 in the sample.
Sample Type Recovery Range(%) Average(%) serum(n=10) 88-104 92 EDTA plasma(n=10) 85-102 94 Heparin plasma(n=10) 88-105 98 - Linearity
-
Dilute the sample with a certain amount of CMLC-1 at 1:2, 1:4 and 1:8 to get the recovery range.
Sample Type 1:2 1:4 1:8 serum(n=10) 85-105% 87-101% 84-92% EDTA plasma(n=10) 88-105% 83-100% 86-99% Heparin plasma(n=10) 94-105% 83-101% 81-100% - Precision(%)
-
Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.
Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.
Item Intra-assay Precision Inter-assay Precision Sample 1 2 3 1 2 3 n 20 20 20 20 20 20 Mean (ng/ml) 1.28 5.03 18.83 1.15 4.87 20.14 Standard deviation 0.07 0.27 0.98 0.06 0.23 0.94 CV(%) 5.23 5.43 5.22 5.12 4.66 4.69 - Stability
-
Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.
ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months Average(%) 80 95-100