Products

Human BIRC7(Baculoviral IAP repeat-containing protein 7) ELISA Kit

Category:

This kit is based on Double antibody-Sandwich ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with anti BIRC7 antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antibody. Then, it binds with BIRC7 bound to precoated antibody. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of BIRC7 in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Alias:BIRC7 ELISA Kit, Livin ELISA Kit, KIAP ELISA Kit, ML-IAP ELISA Kit, baculoviral IAP repeat containing 7 ELISA Kit, baculoviral IAP repeat-containing 7 ELISA Kit, KIAPRING finger protein 50 ELISA Kit, Kidney inhibitor of apoptosis protein ELISA Kit, Livin ELISA Kit, livin inhibitor-of-apoptosis ELISA Kit, Melanoma inhibitor of apoptosis protein ELISA Kit, MLIAPlivin inhibitor of apoptosis ELISA Kit, ML-IAPmLiap ELISA Kit, RNF50LIVINbaculoviral IAP repeat-containing protein 7 ELISA Kit
Catalogue No.:EH2704Species:Human
Range:125-8000pg/mlSensitivity:75pg/ml
  • SPECIFICATIONS
Product Name
Human BIRC7(Baculoviral IAP repeat-containing protein 7) ELISA Kit
Alias
BIRC7 ELISA Kit, Livin ELISA Kit, KIAP ELISA Kit, ML-IAP ELISA Kit, baculoviral IAP repeat containing 7 ELISA Kit, baculoviral IAP repeat-containing 7 ELISA Kit, KIAPRING finger protein 50 ELISA Kit, Kidney inhibitor of apoptosis protein ELISA Kit, Livin ELISA Kit, livin inhibitor-of-apoptosis ELISA Kit, Melanoma inhibitor of apoptosis protein ELISA Kit, MLIAPlivin inhibitor of apoptosis ELISA Kit, ML-IAPmLiap ELISA Kit, RNF50LIVINbaculoviral IAP repeat-containing protein 7 ELISA Kit
Catalogue No.
EH2704
Size
48T/96T
Species
Human
UniProt No.
Q96CA5
Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method
Sandwich ELISA, Double Antibody
Detection Wavelength
OD450
Reaction Duration
4 hours
Range
125-8000pg/ml
Sensitivity
75pg/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with BIRC7 , no obvious cross reaction with other analogues.
ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 Biotin-labeled Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E034 HRP-Streptavidin Conjugate(SABC, 100X) 60ul 120ul
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E049 SABC Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Assay Procedure Summary
Elisa实验原理图
  • Step 1: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate twice without immersing.
  • Step 2: Add 100ul biotin-antibody working solution, seal the plate and statically incubate for 60 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-Streptavidin Conjugate (SABC) working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 0.095 0.097 0.096
125 0.149 0.153 0.151
250 0.212 0.218 0.215
500 0.335 0.345 0.34
1000 0.514 0.528 0.521
2000 0.949 0.977 0.963
4000 1.57 1.616 1.593
8000 2.261 2.327 2.294
Recovery

Add a certain amount of BIRC7 into the sample. Calculate the recovery by comparing the measured value with the expected amount of BIRC7 in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 89-105 96
EDTA plasma(n=10) 85-100 94
Heparin plasma(n=10) 86-101 94
Linearity

Dilute the sample with a certain amount of BIRC7 at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 92-105% 84-95% 88-98%
EDTA plasma(n=10) 85-100% 84-99% 88-97%
Heparin plasma(n=10) 89-105% 89-97% 85-99%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 242 982 3969 257 996 3933
Standard deviation 12.66 48.61 167.89 14.24 46.71 213.56
CV(%) 5.23 4.95 4.23 5.54 4.69 5.43
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100