Human EPRS(glutamyl-prolyl-tRNA synthetase)ELISA Kit

Category:

This kit is based on Double antibody-Sandwich ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with anti EPRS antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antibody. Then, it binds with EPRS bound to precoated antibody. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of EPRS in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
EH8275
Alias: 
EARS ELISA Kit, EPRS ELISA Kit, GLNS ELISA Kit, GLUPRORS ELISA Kit, GluRS Prolyl tRNA synthetase ELISA Kit, Glutamate tRNA ligase ELISA Kit, PARS ELISA Kit, PIG32 ELISA Kit, Proline tRNA ligase ELISA Kit, QARS ELISA Kit, QPRS ELISA Kit
Species: 
Human
Range: 
15.625-1000pg/ml
Sensitivity: 
9.375pg/ml
  • SPECIFICATIONS
Product Name
Human EPRS(glutamyl-prolyl-tRNA synthetase)ELISA Kit
Alias
EARS ELISA Kit, EPRS ELISA Kit, GLNS ELISA Kit, GLUPRORS ELISA Kit, GluRS Prolyl tRNA synthetase ELISA Kit, Glutamate tRNA ligase ELISA Kit, PARS ELISA Kit, PIG32 ELISA Kit, Proline tRNA ligase ELISA Kit, QARS ELISA Kit, QPRS ELISA Kit
Catalogue No.
EH8275
Size
48T/96T
Species
Human
UniProt No.
P07814
Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method
Sandwich ELISA, Double Antibody
Detection Wavelength
OD450
Reaction Duration
4 hours
Range
15.625-1000pg/ml
Sensitivity
9.375pg/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with EPRS , no obvious cross reaction with other analogues.
ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 Biotin-labeled Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E034 HRP-Streptavidin Conjugate(SABC, 100X) 60ul 120ul
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E049 SABC Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Assay Procedure Summary
Elisa实验原理图
  • Step 1: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate twice without immersing.
  • Step 2: Add 100ul biotin-antibody working solution, seal the plate and statically incubate for 60 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-Streptavidin Conjugate (SABC) working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 0.065 0.067 0.066
15.625 0.174 0.18 0.177
31.25 0.285 0.293 0.289
62.5 0.439 0.451 0.445
125 0.823 0.847 0.835
250 1.282 1.32 1.301
500 1.761 1.813 1.787
1000 2.293 2.359 2.326
EH8275 Standard Curve Image
Recovery

Add a certain amount of EPRS into the sample. Calculate the recovery by comparing the measured value with the expected amount of EPRS in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 90-104 98
EDTA plasma(n=10) 90-105 100
Heparin plasma(n=10) 88-103 93
Linearity

Dilute the sample with a certain amount of EPRS at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 92-103% 83-100% 84-92%
EDTA plasma(n=10) 88-96% 82-100% 89-103%
Heparin plasma(n=10) 85-92% 86-98% 90-97%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 31.51 130.4 518 32.84 130 519
Standard deviation 1.75 7.13 21.76 1.82 6.96 28.13
CV(%) 5.56 5.47 4.2 5.53 5.35 5.42
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100