Human C2(Complement C2) ELISA Kit

Manual
  • Specifications
Product name

Human C2(Complement C2) ELISA Kit

Catalogue No.

EH0677

Alias

C2(Complement Component 2) ELISA Kit

Detection method

Sandwich ELISA, Double Antibody

Application

C2 ELISA Kit allows for the in vitro quantitative determination of C2 concentrations in plasma, tissue homogenates and other biological fluids.

Size

96T

Range

0.625-40ng/ml

Sensitivity

< 0.375ng/ml

Standard curve

Human C2(Complement C2) ELISA Kit
Storage

4 ℃ for 6 months

Recovery

Matrices listed below were spiked with certain level of C2 and the recovery rates were calculated by comparing the measured value to the expected amount of C2 in samples.

Matrix Recovery range(%) Average(%)
serum(n=5) 89-101 95
EDTA plasma(n=5) 85-104 95
heparin plasma(n=5) 85-105 95
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of C2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-104% 87-104% 87-103% 87-105%
EDTA plasma(n=5) 82-98% 85-98% 83-101% 87-96%
heparin plasma(n=5) 80-99% 86-93% 83-90% 81-98%
CV(%)

Intra-Assay: CV<8%
Inter-Assay: CV<10%

Note

For Research Use Only

Service

点击这里给我发消息

Hotline

027-87384275
var cnzz_s_tag = document.createElement('script'); cnzz_s_tag.type = 'text/javascript'; cnzz_s_tag.async = true; cnzz_s_tag.charset = 'utf-8'; cnzz_s_tag.src = 'https://w.cnzz.com/c.php?id=1253658600&async=1'; var root_s = document.getElementsByTagName('script')[0]; root_s.parentNode.insertBefore(cnzz_s_tag, root_s);